Model Number |
JINLING-01B |
Specifications Suitable for: Adult, Pediatrics Flow meter: O2, N2O O2 Supple Speed: 35~75L/min Respiratory Rate: 8~40/min I/E Ratio: 1:0.5~1:3.0 Tidal Volume: 100~1200mL (with additional pediatric bellows make tidal volume reach 50~1200ml) Ptr Range: -1.0kPa~1.0kPa Maximum Safe Pressure: ≤6kPa MV Alarm: upper limit: 18L/min lower limit: 0~10L/min Airway Pressure Alarm: upper limit:1~6kPa lower limit: 0.5kPa
Composition & Features ��Anesthesia apparatus: Equipped with 2 vaporizers namely Enflurane and Isoflurane, which can be used alternatively, with temperature control, flow and pressure compensation; 4-tube flow meter; O2-N2O interlock with high precision; integrated circuit system; power cutout protection device; rechargeable battery back up 2 hours after mains supply having been cut. ��Ventilator: Gas driven electrically controlled, time switch, volume constant and pressure restriction; monitoring tidal volume, ventilatory volume, gas-feeding pressure, airway pressure, respiratory frequency, etc. ��Safety device: Alarm upon power having been cut; N2O is cut off automatically if O2 is insufficient; alarm upon the excess of upper & lower airway pressure limit and ventilatory volume. It can be assembled with ECG monitor and Monitor for O2 concentration, and can select vaporizers to assemble
|